A protocol for size-based measurements of nanoplastics across the range 20 nm - 200 nm

AI summary Read the abstract

Researchers validated a method combining asymmetric flow field-flow fractionation with multi-angle light scattering to size and quantify polystyrene nanoplastics in the 20–200 nanometre range. The study found that smaller nanoplastics are systematically underestimated by the technique and that separation conditions need optimisation for accurate measurement.

Polymers

The on-line coupling of asymmetric flow field-flow fractionation (AF4-FFF) to multi-angle light scattering (MALS) and UV diode array (UV-DAD) detectors to simultaneously separate, collect information about the size and quantify the polystyrene nanoplastics (PS-NPs) was tested. A mixture of PS-NPs with nominal size of 20±2 nm, 60±4 nm, 100±3 nm and 200±5 nm was prepared by dilution with ultrapure deionized water and gentle mixing. From the AF4- FFF results, the smaller nanoplastics (20 nm) had a dominant peak at short elution time while the larger ones (200 nm) had the dominant peak at high elution time, and nanoplastics at 60 nm and 100 nm eluted in the middle region. From MALS analysis using the 90° scattering angle, the obtained size of the different peaks was consistent with the values provided by the supplier. By UV-DAD detection, concentrations of the nanoplastic populations were measured at the wavelength of 215 nm. The power of AF4-FFF resides in its essential tunability: simply by changing flow rates, a single separation channel can be used to explore complex samples across the size range 20 nm-200 nm, with high resolution between differently sized nanoplastics, and providing the possibility to identify sample components through coupling to on-line detectors.

More Papers Like This

Article Tier 2

Validated method for polystyrene nanoplastic separation in aqueous matrices by asymmetric-flow field flow fraction coupled to MALS and UV–Vis detectors

AI summary Read the abstract

Researchers developed and fully validated a method to accurately measure nanoplastic particle sizes (30–490 nm) in water using a technique that combines flow separation with light-scattering detection. Having a validated analytical method is a critical step for standardizing how nanoplastics are measured across laboratories, enabling more consistent assessment of their environmental risks.

Article Tier 2

In-house validation of AF4-MALS-UV for polystyrene nanoplastic analysis

AI summary Read the abstract

Researchers validated an asymmetric flow field-flow fractionation method coupled with multi-angle light scattering and UV detection for simultaneously sizing and quantifying polystyrene nanoplastics from 20 to 200 nm, achieving greater than 90% recovery and detection limits of 15–33 µg/mL across a polydisperse nanoplastic mixture.

Article Tier 2

A method for efficient separation of polystyrene nanoplastics and its application in natural freshwater

AI summary Read the abstract

Researchers developed a method using asymmetrical flow field-flow fractionation (AF4) coupled with multiple detectors to efficiently separate and characterize polystyrene nanoplastics by particle size in freshwater environments, demonstrating its applicability for analysing nanoplastic environmental behaviour in natural freshwater samples.

Article Tier 2

Detection of nanoplastics in food by asymmetric flow field-flow fractionation coupled to multi-angle light scattering: possibilities, challenges and analytical limitations

AI summary Read the abstract

Researchers tested whether asymmetric flow field-flow fractionation coupled to light-scattering detection could identify nanoplastics in fish tissue, successfully detecting 100 nm polystyrene particles at a limit of 52 µg/g after enzymatic digestion, but found the method could not be directly transferred to polyethylene particles without further adjustment.

Article Tier 2

An effective solution to simultaneously analyze size, mass and number concentration of polydisperse nanoplastics in a biological matrix: asymmetrical flow field fractionation coupled with a diode array detector and multiangle light scattering

AI summary Read the abstract

Researchers developed an asymmetrical flow field-flow fractionation method coupled with a diode array detector and multiangle light scattering to simultaneously measure the size, mass, and number concentration of polydisperse nanoplastics in biological matrices, providing a more accurate tool for assessing nanoplastic pollution levels.