In-house validation of AF4-MALS-UV for polystyrene nanoplastic analysis
AI summary Read the abstract
Researchers validated an asymmetric flow field-flow fractionation method coupled with multi-angle light scattering and UV detection for simultaneously sizing and quantifying polystyrene nanoplastics from 20 to 200 nm, achieving greater than 90% recovery and detection limits of 15–33 µg/mL across a polydisperse nanoplastic mixture.
The suitability of asymmetric flow field-flow fractionation (AF4) coupled on-line to multi-angle light scattering (MALS) and UV diode array (UV-DAD) detectors was tested to simultaneously detect polystyrene nanoplastics (PS-NPs) and collect information about their size. A mixture of four sizes of PS-NPs at 20 nm, 60 nm, 100 nm and 200 nm was prepared by dilution with ultrapure deionized water and gentle mixing and was used as test sample for a polydisperse nanoplastic system. The AF4 method separated each single size of PS-NP mixture in a total time of 48 min by using 0.2% SDS as carrier solution. Then, the PS-NPs were sized and detected by following their MALS (90° scattering angle) and UV (215 nm) signals. Quality control (QC) performances as linearity, between-day repeatability, resolution factor, trueness/recovery, limit of detection (LoD) and selectivity were calculated, according to the ISO/TS 21362:2018. Method uncertainty was calculated following the ISO/TS 21748:2002 by summing between-day repeatability and trueness or recovery uncertainties. In-house validation results demonstrated good peak resolution and selectivity, R linearity of 0.998-0.999 in the range 50-1000 μg/mL, between-day repeatability of ca. 10%, trueness/recovery above 90% and LoD between 15 μg/mL (20 nm) and 33 μg/mL (200 nm). Expanded uncertainty was 16.1-17.9% on PS-NP size between 60 and 200 nm and 10.4-14.7% on PS-NP concentration between 100 and 1000 μg/mL. Compared to traditional single-technique analysis, this hyphenated method offers great promise for separating and analysing diverse populations of PS-NPs present in real matrices, which is critical for health and risk assessment studies and any regulatory action.
More Papers Like This
A protocol for size-based measurements of nanoplastics across the range 20 nm - 200 nm
AI summary Read the abstract
Researchers validated a method combining asymmetric flow field-flow fractionation with multi-angle light scattering to size and quantify polystyrene nanoplastics in the 20–200 nanometre range. The study found that smaller nanoplastics are systematically underestimated by the technique and that separation conditions need optimisation for accurate measurement.
Validated method for polystyrene nanoplastic separation in aqueous matrices by asymmetric-flow field flow fraction coupled to MALS and UV–Vis detectors
AI summary Read the abstract
Researchers developed and fully validated a method to accurately measure nanoplastic particle sizes (30–490 nm) in water using a technique that combines flow separation with light-scattering detection. Having a validated analytical method is a critical step for standardizing how nanoplastics are measured across laboratories, enabling more consistent assessment of their environmental risks.
A method for efficient separation of polystyrene nanoplastics and its application in natural freshwater
AI summary Read the abstract
Researchers developed a method using asymmetrical flow field-flow fractionation (AF4) coupled with multiple detectors to efficiently separate and characterize polystyrene nanoplastics by particle size in freshwater environments, demonstrating its applicability for analysing nanoplastic environmental behaviour in natural freshwater samples.
Detection and Identification of Non-Labeled Polystyrene Nanoplastics in Rodent Tissues Using Asymmetric Flow Field-Flow Fractionation (AF4) Combined with UV–Vis, Dynamic Light Scattering (DLS) Detectors and Offline Pyrolysis–GCMS (Pyro-GCMS)
AI summary Read the abstract
Researchers developed a multi-platform analytical method combining field-flow fractionation, light scattering, and pyrolysis-based mass spectrometry to detect and quantify non-labeled polystyrene nanoplastics in rodent tissues. They created tissue-specific digestion protocols for intestine, kidney, and liver that maximize particle recovery while preventing loss of submicron particles. The method provides a standardized approach for tracking nanoplastics in biological samples, which is essential for evaluating potential health risks.
New method for separating and online detecting polydisperse mixed nanoplastics
AI summary Read the abstract
Researchers optimized an asymmetric flow field-flow fractionation method coupled with UV detection to separate and quantify mixed nanoplastics between 20 and 200 nm in a single run, achieving high recovery rates and low detection limits across multiple polymer types and real water sample conditions.
Research digests by email
When a large batch of papers lands in the Atlas, we read through it and send a short write-up of what stood out.