Pre-analytical challenges from adsorptive losses associated with thiamine analysis
AI summary Read the abstract
Researchers investigating thiamine (vitamin B1) analysis found that commonly used glass fiber filters and glass vials cause dramatic analyte losses—recovering as little as 1% of a 100 nM standard—and recommend switching to polymeric filters and polypropylene or polycarbonate storage containers to ensure accurate measurement of this essential vitamin.
Thiamine (vitamin B1) is an essential vitamin serving in its diphosphate form as a cofactor for enzymes in the citric acid cycle and pentose-phosphate pathways. Its concentration reported in the pM and nM range in environmental and clinical analyses prompted our consideration of the components used in pre-analytical processing, including the selection of filters, filter apparatuses, and sample vials. The seemingly innocuous use of glass fiber filters, glass filter flasks, and glass vials, ubiquitous in laboratory analysis of clinical and environmental samples, led to marked thiamine losses. 19.3 nM thiamine was recovered from a 100 nM standard following storage in glass autosampler vials and only 1 nM of thiamine was obtained in the filtrate of a 100 nM thiamine stock passed through a borosilicate glass fiber filter. We further observed a significant shift towards phosphorylated derivatives of thiamine when an equimolar mixture of thiamine, thiamine monophosphate, and thiamine diphosphate was stored in glass (most notably non-silanized glass, where a reduction of 54% of the thiamine peak area was observed) versus polypropylene autosampler vials. The selective losses of thiamine could lead to errors in interpreting the distribution of phosphorylated species in samples. Further, some loss of phosphorylated thiamine derivatives selectively to amber glass vials was observed relative to other glass vials. Our results suggest the use of polymeric filters (including nylon and cellulose acetate) and storage container materials (including polycarbonate and polypropylene) for thiamine handling. Losses to cellulose nitrate and polyethersulfone filters were far less substantial than to glass fiber filters, but were still notable given the low concentrations expected in samples. Thiamine losses were negated when thiamine was stored diluted in trichloroacetic acid or as thiochrome formed in situ, both of which are common practices, but not ubiquitous, in thiamine sample preparation.
More Papers Like This
Examination of Microcystin Adsorption by the Type of Plastic Materials Used during the Procedure of Microcystin Analysis
AI summary Read the abstract
Researchers tested how well different plastic container and filter materials retain microcystins during sample preparation for water quality analysis, finding that polyethylene terephthalate containers had the best recovery rate (at least 81%) over 48 hours. The results show that plastic material choice in laboratory procedures can significantly affect the accuracy of microcystin measurements in drinking water.
Potential sources of microplastic contamination in laboratory analysis and a protocol for minimising contamination
AI summary Read the abstract
Scientists identified multiple sources of microplastic contamination that can be accidentally introduced during laboratory analysis of biological samples, including from lab water, chemicals, and air. By filtering reagents, pre-treating glass fiber filters, and using a clean working environment, they reduced lab-introduced contamination by 70–100%. These findings are important for accurate microplastic research: without such precautions, studies can significantly overestimate how much plastic is actually present in animal tissues or environmental samples.
Assessing the Impact of Shipping on Microplastic Concentration of Filtered Samples
AI summary Read the abstract
Researchers evaluated the influence of different shipping and packaging methods on microplastic recovery rates from metal filters, using water samples spiked with polyethylene spheres to assess whether filtering and mailing samples in metal tins could serve as a standardized transport method for microplastics research.
Evaluation of filter substrates for concurrent optical photothermal infrared and Raman spectroscopy in microplastic analysis
AI summary Read the abstract
Scientists testing tiny plastic particles need reliable tools, and this study found that the material used to hold samples (the "filter") can make a big difference in how accurately advanced scanning techniques detect microplastics. By testing nine different filter types, researchers identified two (Anodisc and gold-coated PET) that gave the clearest, most consistent results, helping labs choose better tools to detect the tiny plastic particles increasingly found in our food, water, and even bodies. This matters because more accurate detection methods are the first step toward understanding how much microplastic exposure humans actually face and what health risks it might pose.
Identifying laboratory sources of microplastic and nanoplastic contamination from the air, water, and consumables
AI summary Read the abstract
Researchers systematically tested common laboratory materials and procedures for plastic contamination and found that water sources, airflow, dust, and even glassware can introduce significant amounts of micro- and nanoplastics into samples. Surprisingly, glass consumables and aluminum foil, both commonly recommended as contamination controls, were themselves sources of plastic particles. The findings highlight that current best practices for preventing sample contamination in microplastic research may need to be revised.