0
Article ? AI-assigned paper type based on the abstract. Classification may not be perfect — flag errors using the feedback button. Tier 2 ? Original research — experimental, observational, or case-control study. Direct primary evidence. Marine & Wildlife Policy & Risk Sign in to save

Compact off-axis holographic slide microscope: design guidelines

Biomedical Optics Express 2020 46 citations ? Citation count from OpenAlex, updated daily. May differ slightly from the publisher's own count.
Teresa Cacace, Vittorio Bianco, Biagio Mandracchia, Vito Pagliarulo, Emilia Oleandro, Melania Paturzo, Pietro Ferraro

Summary

Researchers designed a compact, 3D-printed off-axis holographic microscope that can identify and image microparticles in fluid in real time. This portable, low-cost device could be deployed in the field for monitoring microplastic contamination in water samples.

Polymers

Holographic microscopes are emerging as suitable tools for in situ diagnostics and environmental monitoring, providing high-throughput, label-free, quantitative imaging capabilities through small and compact devices. In-line holographic microscopes can be realized at contained costs, trading off complexity in the phase retrieval process and being limited to sparse samples. Here we present a 3D printed, cost effective and field portable off-axis holographic microscope based on the concept of holographic microfluidic slide. Our scheme removes complexity from the reconstruction process, as phase retrieval is non iterative and obtainable by hologram demodulation. The configuration we introduce ensures flexibility in the definition of the optical scheme, exploitable to realize modular devices with different features. We discuss trade-offs and design rules of thumb to follow for developing DH microscopes based on the proposed solution. Using our prototype, we image flowing marine microalgae, polystyrene beads, E.coli bacteria and microplastics. We detail the effect on the performance and costs of each parameter, design, and hardware choice, guiding readers toward the realization of optimized devices that can be employed out of the lab by non-expert users for point of care testing.

Share this paper