We can't find the internet
Attempting to reconnect
Something went wrong!
Hang in there while we get back on track
Characterization in respect to degradation of titanium‐coated polypropylene surgical mesh explanted from humans
Summary
Researchers analyzed titanium-coated polypropylene surgical mesh samples explanted from human patients to assess whether the titanium coating protects against degradation. Using multiple analytical techniques including infrared spectroscopy and electron microscopy, the study examined how the mesh materials changed during clinical use, providing the first characterization of residual material properties in Ti-coated mesh post-implantation.
Titanium-coated polypropylene (Ti-PP) mesh was introduced in 2002 as a surgical mesh for the treatment of hernias and shortly after for pelvic floor surgery, with the aim of improving biocompatibility when compared to non-titanised/regular PP mesh implants. The application of a titanium coating could also be beneficial to address concerns regarding the exposure of PP in an in vivo environment. Many studies have shown that PP, although it is widely accepted as a stable polymer, is subject to oxidation and degradation, such degradation affects the mechanical behavior, that is, the stiffness and tensile strength of PP mesh. Despite the wide clinical use of Ti-PP surgical meshes, no study has yet investigated the residual material properties post clinical deployment and subsequent explantation. In this study, two explanted Ti-PP mesh samples each having different incorporation durations from two patients were examined. Material analysis conducted within this study includes the following techniques: attenuated total reflectance-Fourier transform infrared spectroscopy (ATR-FTIR), Raman spectroscopy, low voltage - scanning electron microscopy (LV-SEM), backscattered electron (BSE) imaging, energy dispersive X-ray spectroscopy (EDS) and secondary election hyperspectral imaging (SEHI). The hypothesis of this study is that the Ti coating successfully shields the PP mesh from oxidative stress in vivo and thus protects it from degradation. The results of this analysis show for the first time evidence of bulk oxidation, surface degradation, and environmental stress cracking on explanted Ti-PP meshes.
Sign in to start a discussion.
More Papers Like This
Influence of polypropylene mesh degradation on tissue inflammatory reaction
Polypropylene surgical mesh implants were found to degrade in vivo through surface cracking and peeling, releasing particles and altering the inflammatory tissue response around the implant. The study raises concerns about the long-term biocompatibility of polypropylene mesh given its degradation behavior.
Characterization and quantification of oxidative stress induced particle debris from polypropylene surgical mesh
Researchers characterized and quantified particle debris released from polypropylene surgical mesh under oxidative stress conditions, raising concerns about microplastic generation from implanted medical devices and potential host tissue responses.
Photocatalytic aging process of Nano-TiO2 coated polypropylene microplastics: Combining atomic force microscopy and infrared spectroscopy (AFM-IR) for nanoscale chemical characterization
Researchers investigated the photocatalytic aging of nano-TiO2 coated polypropylene microplastics using AFM-IR nanoscale chemical characterization, finding that the TiO2 coating accelerated surface degradation and altered the physicochemical properties of the microplastics compared to pristine and conventionally aged counterparts.
Corrosion and Electrochemical Impedance Spectroscopy of Thin TiALN and TiCN PVD Coatings for Protection of Ballast Water Screen Filters
Researchers evaluated the corrosion resistance of titanium-based coatings applied to stainless steel mesh under simulated marine conditions. Corrosion-resistant materials are relevant to the design of durable marine monitoring equipment, including tools used to sample and filter microplastics from seawater.
Uncovering the relationship between macrophages and polypropylene surgical mesh
Researchers studied how polypropylene surgical mesh, commonly used to treat stress urinary incontinence, degrades inside the body when exposed to mechanical forces and oxidation. The degraded mesh surface cracked and triggered inflammatory and scarring responses in immune cells, helping explain why these implants sometimes cause serious complications. This study demonstrates that plastic materials can break down within the body and provoke harmful immune responses, which is directly relevant to concerns about microplastic accumulation in human tissues.